a) If the Interlock indicator on the Station control panel is active, click the Reset button to clear the
interlock.
b) Apply power to the hydraulic service manifold (HSM) for the test station by clicking the HSM Low button
and then the High button on the Power control panel.
4.2.0 Install a Specimen
4.2.1.0 Before You Begin
Before you can install a specimen, you must open a test station and apply hydraulics. For more information,
see Starting Station Manager and Loading Station Configuration on page 18 and Starting MPX and Applying
Hydraulic Power on page 19.
4.2.2.0 Setting Station Limits in Station Manager Application
To reduce the risk of injury when installing specimens, set station limits that limit the force applied by the
actuator. After installing the specimen, you may have to adjust these limits to allow the test to run.
Note:
For more information about setting limits, see MTS Series 793 Control Software.
1. Maximize the Station Manager window.
Warning:
Actuators can produce dangerous forces.
A moving actuator can injure anyone in its path.
Always set station force limits before installing a specimen.
2. Set the station limits and the resulting action.
20 MTS TestSuite
4.0 Operator Procedures