SWIFT 40 Sensors
66
Edit the Calibration File
Setting up the Transducer Interface
2. If necessary, edit the value for OutputPolarities.
This value sets the polarity states of all six outputs (Mz, My, Mx, Fz, Fy, and
Fx) using the following binary scheme (the default configuration is
represented):
For example, if you want to invert Fx and Fy, and Mx and My, (making their
polarities opposite those shown on the labeling) you would make the
following changes:
O
UTPUTS MZ MY MX FZ FY FX
Bit
2
5
2
4
2
3
2
2
2
1
2
0
Bit Value 32 16 8 4 2 1
Example 101000
= 40
10
(default)
M
Z MY MX FZ FY FX
Bit
2
5
2
4
2
3
2
2
2
1
2
0
Bit Value 32 16 8 4 2 1
101000= 40
10
(old value)
110011
= 51
10
(new value)
*
* See the examples in the next table.