Error Messages
SWIFT
®
Mini TI
Software Utilities
39
Zero
Good – no problems detected
Bad Angle Offset – The change in angle was not 90°, ±2°.
Direction Change – The direction of 90° increments reversed.
Shunt
Good – no problems were detected.
Reference Bad The shunt reference is out of range.
Shunted Bad – A shunt value does not match the shunt reference.
Unshunted Bad – While shunting one bridge another bridge output
unexpectedly changed.
Calibration Data is:
Good – No problems were detected.
Bad – The unit has not been calibrated or the calibration memory is
corrupted.