MTS MultiPurpose TestWare®
Set Process Parameters in the Group Table
Creating a Practice Test Procedure
172
Set Process Parameters in the Group Table
1. Display the procedure table that pertains to the group process.
A. Click the Group process icon on the MPT procedure table.
B. Click on the MPT Procedure Editor toolbar.
2. Define the Data Limit Detector process (Trigger at 75C).
A. In the group table, click the Data Limit Detector (Trigger at 75C)
process icon.
B. Click the Signals tab. Move the Temperature Monitor to the Included
list and ensure that all other signals are in the Available list.
C. Click the Limits tab. Type and select the parameters needed for the
test, as follows:
Signal: Temperature Monitor
Upper Limit (enabled): 75.0 degrees C.
Lower Limit (disabled)
D. Observe the default settings on the Settings, Limit Table, and General
tabs.
Data Limit Detect
(Trigger at 75C)