Concentric Adjustments
709 Alignment System Alignment Wizard
113
If the upper gauge Y coordinate is positive
1. Initially loosen C0.
2. Tighten C180 moving the red (upper) and blue (lower) gauge readings
together until they slightly pass each other.
3. Tighten C0 to bring the red and blue gauge readings back to
horizontally align with each other as close as practical.