
Routine Maintenance Overview Checklist
Important:
There are no customer serviceable components on the MTS Acumen frames. Maintenance consists
of keeping the frame and work area clean, general inspection, checking interlocks, and scheduled
frame calibration.
Recommended Service
AnnuallyMonthlyDailyCalendar Time Using 8 Hours Running
Time Rate per Day
20001608Running Time-Hours
XClean work area / Machine Surfaces
XCheck Frame Mounted Control Emergency
Stop
XCheck Remote Emergency Stop (If
Present)
XCheck Test Area Enclosure - Open Door
Applies Low Velocity
Load Frame / Crosshead
MTSCheck Crosshead Locks For Proper
Operation
MTSCheck Crosshead Lift Belt For Wear
MTSCheck / Adjust Crosshead Lift Belt Tension
MTSCheck / Clean Crosshead Lift Screws
MTSCheck / Clean Columns
MTSAssure Crosshead Movement Is Smooth
Actuator
MTSCheck Cooling System Is Operational
MTSClean Exposed Cooling Surfaces
MTSAssure Actuator Movement Is Smooth
MTSCheck Actuator Guidance Mechanism
MTSCheck Brake Operation
62 MTS Acumen™ Electrodynamic Test System
Maintenance
Kommentare zu diesen Handbüchern