Model 493.74 Two-Station HSM Interface Board
MTS FlexTest® Models 40/60/100/200 Controller Hardware Digital I/O and Transition Boards
253
Cable specification To maintain EMC compliance, J43 Interlock cables must comply with the
following specifications:
Connector–9-contact, type D, male EMI connector.
Backshell–EMI metallized plastic or metal.
Cable–AWG and number of conductors as required. Braided shield with
shield connected to the metallized backshell at the chassis.
Jumper plug If connector J43 is not used, you must install a jumper plug to maintain the
integrity of the interlocks. Use jumper plug part number 100-007-948 or
jumper pins 1 and 2; 3 and 4; 5 and 6; 7 and 8.