Mts Multipurpose Elite Bedienungsanleitung Seite 404

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 662
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 403
Sensitivity
A sensitivity value specifies how much the signal must change to detect a peak or valley. It can be used
to keep the activity from seeing signal noise as a new peak or valley:
Setting the sensitivity too low may cause signal noise to be recognized as peaks and valleys.
Setting the sensitivity too high may cause low amplitude signals to be missed.
Peak-Valley Change Detector Sensitivity
Sensitivity Properties
DescriptionCallout
SensitivityA
Detected PeaksB
Detected ValleysC
Tolerance
The tolerance band specifies how much a peak or valley must deviate from the associated reference levels
before it generates a trigger signal.
For example, if the established peak value is +5 cm, and the tolerance is set to ±0.25 cm, the activity does
not generate a trigger signal unless the peak value registers greater than +5.25 cm or less than +4.75 cm.
Peak-Valley Change Detector Tolerance
Tolerance Properties
DescriptionCallout
PeakA
404 | MTS TestSuite
Test Activities
Seitenansicht 403
1 2 ... 399 400 401 402 403 404 405 406 407 408 409 ... 661 662

Kommentare zu diesen Handbüchern

Keine Kommentare