How to Provide Program to a Series 407 Controller
Models FlexTest® IIm/GT/SE Controller Hardware FlexTest GT Controller Connections
155
D. On the 407 Controller, clear the hydraulic Interlock:
–Press the Alt Function button.
–Press the Reset button (“9” on the key pad). The Interlock indicator
on the front panel should extinguish.
E. On the 407 Controller, apply hydraulic pressure with the Hydraulic
Pressure buttons.
F. On the 407 Controller, press Run. This step readies the 407 Controller
to receive program from the 793 Controller.
G. On the Station Manager application’s Station Controls panel, click
Reset to clear the Program interlock.
H. On the Station Manager application’s Station Controls panel, click the
(Program Run). This step sends program from the 793 Controller
to the 407 Controller.