Models FlexTest® IIm/GT/SE Controller Hardware
Model 493.02 Chassis Multiple Controller
Hydraulic Configurations
246
Model 493.02 Chassis Multiple Controller Interconnections
FlexTest SE Controllers support the following hydraulic interconnections for
multiple controller configurations:
• Independent HSM control with no HPU
• Independent HSM control with a shared HPU
• Shared HSM with an HPU
• Shared HSM without HPU
• Independent HSM controlled with a shared HPU using first-on-last-off
For more detailed information about multiple controller configurations,
see“Multiple Controller Connections” on page 220 and the FlexTest SE User
Manual.
Independent HSM (no HPU)
For this multiple controller configuration:
• Each controller has an independent hydraulic service manifold (HSM).
• No hydraulic power unit (HPU) is connected.
• Select HSM Only for Power Type on each connected controller.
Path: Config > Edit Hardware > Power Options > Power Type > HSM
Only
FlexTest SE
Emergency Stop
7
8 9
65
4
3
2
1
.
0
+/-
enter
cancel
FG
Setup
Status
Scope
Meters
Tuning
Config
Limits
Log
Menu
?
A
B
Recall
Enabled
reset
Rewind Run
Hold
Stop
Power
Program
Hydraulic
Interlock
Program
Interlock
HPU
HSM
Off
Low
High
Off
Low
High
Navigate
Monitor 1
Monitor 2
FlexTest SE
Emergency Stop
7
8 9
65
4
3
2
1
.
0
+/-
enter
cancel
FG
Setup
Status
Scope
Meters
Tuning
Config
Limits
Log
Menu
?
A
B
Recall
Enabled
reset
Rewind Run
Hold
Stop
Power
Program
Hydraulic
Interlock
Program
Interlock
HPU
HSM
Off
Low
High
Off
Low
High
Navigate
Monitor 1
Monitor 2
Power Type = HSM Only
Power Type = HSM Only
Interlock/Hydraulics
HSM 1
056-126-4xx
J28 HSM
HSM 2
056-126-4xx
J28 HSM
056-316-0xx
Kommentare zu diesen Handbüchern