Mts TestLine Bedienungsanleitung Seite 63

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 132
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 62
Leave the Operator Type selection set to Tuning because you will do more tuning once the specimen
is installed.
Power Up the Pump and HSM
HPU setups vary considerably by lab. This document describes a typical HPU setup. For information on
your particular system, consult your lab experts or system documentation.
1. Correct any interlock conditions.
If a displacement limit has been exceeded, see If Necessary, Recover from a Tripped Displacement
Limit (p. 113).
2. Reset the interlock.
Click the Reset button. If the interlock will not reset, there is likely an interlock condition that has not
been corrected.
3. Open the handvalve.
4. Power up the HPU (hydraulic pump unit) and HSM (hydraulic service manifold).
MTS TestLine Test System Operation | 63
Running the Example HCF Test
Seitenansicht 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 131 132

Kommentare zu diesen Handbüchern

Keine Kommentare