Use the following parameters:
• Meter Type — Timed
• Channel — Ch1 (or other depending on your configuration)
• Signal — Force
• Engineering Units — kN
Zero Displacement and Force Signals
With the actuator at mid-travel and no force on the load cell, zero the displacement and force signals.
1. Auto offset the Displacement signal.
a) Click the Station Setup button.
b) Select Channels > Ch1 > Displacement > Channel Input Signals button > Offset/Zero tab.
c) Click the Auto Offset button.
2. Auto offset the Force signal.
80 | MTS TestLine Test System Operation
Running the Example HCF Test